4-6 System Configuration 879596 TempScan / MultiScan User's Manual
Required Configuration
Although most of the user-defined configuration states have associated defaults, the TempScan/1100 or
MultiScan/1200 does not scan its channels on the initial power-up activation. Before scans can begin, the
following required configurations must be performed:
Configuration of the desired channels. See command Channel Configuration (
C).
Configuration of the scan. See commands Channel Configuration (
C) and Set Scan Interval (I).
Configuration of the acquisition. See command Set Counts (
Y).
Configuration of the Trigger and Stop events. See command Set Trigger Configuration (
T).
After these steps are completed and scanning begins, the alarms and the High/Low/Last Registers are
updated at the programmed scan rate, and the scans are placed in the Acquisition Buffer.
Example 12a. Required Configurations
(1) PRINT#1,"OUTPUT07; C1-4,1X"
(2) PRINT#1,"OUTPUT07;I00:01:00.0,00:00:01.0X"
(3) PRINT#1,"OUTPUT07;Y100,1000,0X"
(4) PRINT#1,"OUTPUT07;T1,8,0,0X"
(5) PRINT#1,"OUTPUT07;@X"
(6) PRINT#1,"OUTPUT07 U4X"
(7) PRINT#1,"ENTER07"
(8) INPUT A$
The above command lines illustrates all of the required configurations:
Line 1: Configure channels 1 to 4 with Type J thermocouples.
Line 2: Configure the normal scan interval to 1 minute and the acquisition scan interval to 1 second.
Line 3: Set the Pre-Trigger scan count to 100 and the Post-Trigger scan count to 1000.
Line 4: Configure the Trigger (trigger start event) to the software Trigger On (
@) command and the
Stop (trigger stop event) to the Post-Trigger Count. The front panel TRIGGER LED indicator should
flash.
Line 5: Trigger the acquisition. Acquisition scans should now be collected until 1000 Post-Trigger
scans have been collected. The front panel TRIGGER LED indicator should be on.
Querying the High/Low/Last Registers shows the temperature values for the configured channels.
Line 6: Query HLL registers for the configured channels.
Line 7: Get the response.
Line 8: The screen will show the temperature data for the four channels.