Measurement-computing TempScan/1100 Bedienungsanleitung Seite 120

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 306
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 119
6-4 System Calibration 879596 TempScan / MultiScan User's Manual
Calibration of Master Chassis
Note: The following text discusses the manual method of calibration. However, we strongly
suggest using ChartView’s internal automated calibration method. To do so, from
ChartView navigate as follows, then follow the on screen instructions.
Device Select Status Calibration [then follow on screen prompts]
Note: Prior to each calibration session, disable calibration protection by setting the rear panel
DIP microswitch 9 to the up (1) position. Then enable Calibration Mode via the Enter
Calibration Mode (K) command with a valid password. If the default calibration
password has not been changed, enter K12345.
Note: At the end of each calibration session, disable Calibration Mode with the End Calibration
Mode (E) command. Then re-enable calibration protection by resetting DIP
microswitch 9 to the down (0) position.
Seitenansicht 119
1 2 ... 115 116 117 118 119 120 121 122 123 124 125 ... 305 306

Kommentare zu diesen Handbüchern

Keine Kommentare