Measurement-computing TempScan/1100 Bedienungsanleitung Seite 126

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 306
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 125
6-10 System Calibration 879596 TempScan / MultiScan User's Manual
To Calibrate a Thermocouple Card for Cold-Junction Offset
CAUTION
Ensure that the thermocouple scanning card has been calibrated for offset and
gain. If not, follow the thermocouple card offset and gain calibration procedures
as described in the previous text.
Cold-Junction Offset Calibration
1. Connect accurate thermocouple sources to the following multiple channels on the scanning card being
calibrated, where
chan refers to the first channel number on the scanning card being calibrated (refer
to the previous table for valid cold-junction channel numbers).
TempTC/32B (for TempScan/1100): chan, chan+4, chan+16, chan+20
TempTC/32A (for TempScan/1000A): chan, chan+4, chan+16, chan+20
MTC/24 (for MultiScan/1200): chan, chan+8, chan+16
Slide the scanning card into the previously-calibrated master chassis.
2. Power up the unit and wait at least one hour before continuing.
3. Apply the following temperature values in Celsius to the multiple cold-junction channels above:
TempTC/32B (for TempScan/1100): nnn.n, ppp.p, qqq.q, rrr.r
TempTC/32A (for TempScan/1000A): nnn.n, ppp.p, qqq.q, rrr.r
MTC/24 (for MultiScan/1200): nnn.n, ppp.p, qqq.q