Measurement-computing Analyzer488 Bedienungsanleitung Seite 60

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 281
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 59
Section 3 Front Panel Operation
3.14
3.7.4 Viewing Recorded Bus Events
To look at the events that were just recorded,
Press
[VIEW]
The
VIEW
and
MEMORY
LEDs will light
SRQ IFC
RENATN
EOI
LOCATION MESSAGE
00000 TAG21 55
¼
¼
The first controller command,
TAG21
, is stored in record memory
location
0
.
SRQ IFC
RENATN
EOI
LOCATION MESSAGE
Press
[
]
eight times
00001 UNL 3F
¼
¼
to see the rest:
00002 LAG00 20
¼
¼
00003 't' 74
¼
¼
00004 'e' 65
¼
00005 's' 73
¼
00006 't' 74
¼
00007 CR OD
¼
¼
00008 empty - - - -
3.7.5 Using Triggers
The following example shows how to set up a trigger, record some
bus data, view the record memory, search the record memory, and
review recording statistics. To set up a trigger point, press
[SETUP]
to configure the trigger parameters:
SRQ IFC
RENATN
EOI
LOCATION MESSAGE
Press
[
][2][0][
][1]
Match? LAG00 20X1XXX
[ENTER]
to match
&h20
(LAG00)
Seitenansicht 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 280 281

Kommentare zu diesen Handbüchern

Keine Kommentare