Section 4 Serial Controller
4.20
TRIGGER
The event message '
TRIGGER
' is returned to the serial host when the
Analyzer488 trigger circuitry has found the trigger point. This event
is edge sensitive.
RECORD
The event message '
RECORD
' is returned to the serial host when the
Analyzer488 recording sequence is finished. This event is edge
sensitive.
STEP
The event message '
STEP
' is returned to the serial host when the bus
is ready for a source (talker) or acceptor (listener) handshake. This
event is edge sensitive.
MODE
any
BUS STATES None
QUERY String of events in the following format:
RESPONSE ARM SRQ TRIGGER STEP RECORD
EXAMPLES
100 PRINT#1,"ARM SRQ"
Enable SRQ Condition
110 T = TIMER
110 IF LOC(1) THEN PRINT INPUT$(LOC(1),1);
120 IF TIMER - T < 5 GOTO 110
SRQ message will be received and printed if an SRQ occurs within
5 seconds after receipt of the
ARM
command.
PRINT #1,"ARM SRQ STEP"
Arm SRQ and STEP.
PRINT #1,"ARM?"
Query armed status.
INPUT #1, A$
Get response.
PRINT A$
Display results
The returned string would be:
'ARM SRQ STEP'
Refer to the demonstration program listed in appendix F for further examples of
the ARM command.