Table 7-1. User-Specified Gain Resistor Positions
RX1077RX1033
RX1066RX1022
RX1055RX1011
RX1044RX1000
Resistor PositionChannelResistor PositionChannel
10 Ohms800
17 Ohms700
40 Ohms500
130 Ohms300
161 Ohms200
364 Ohms100
Resistor ValueGain
The equation for selecting the gain resistor, R
USER,
for any gain between X100 and X1000 is:
R
USER
= (40000 / (Gain 1) ) 40
7.7 Setting the Input Configuration
CHANNEL CONFIGURATION SWITCHES SET
VOLTAGE, THERMOCOUPLES, OR 2/4-WIRE RTDs
BOTH 4s ARE ON (DOWN), BOTH 3s ARE OFF (UP)
3
4
3
4
CH0
O
N
O
N
CH4
IN CONFIG
3
4
3
4
Channel Configuration Switch - Voltages
A channel configuration switch is associated with each
channel (Figure 7-5). The switches are used to configure the
input circuits for voltage inputs, thermocouple inputs, 2, 3, or
4-wire RTDs and bridges.
For bridge measurements on a particular channel, set the
switches labeled “4” to the ON (down) position for that
channel.
Set the switches labeled “3” in the OFF (up) position.
Figure 7-5.
Input Channel Configuration Switches
7.8 Configuring the Bridge
As mentioned earlier in this chapter, resistance measurements are made by constructing a bridge
containing precision resistors with known values against which the unknown resistor is to be compared.
In strain gauge applications, the strain gauge sensor itself may make up a quarter of this bridge, half of
30