18
Chapter 5
Specifications
All specifications are subject to change without notice.
Typical for 25°C unless otherwise specified.
Specifications in italic text are guaranteed by design.
Analog input
Table 1. Generic analog input specifications
Four dual 24-bit, Sigma-Delta type
500 VDC minimum between field wiring and
USB interface
Differential input voltage
range
Absolute maximum input
voltage
±C0x through ±C7x relative to GND
(pins 9, 19, 28, 38)
±25 V power on, ±40 V power off
Open thermocouple detect enabled
Normal mode rejection ratio
Common mode rejection ratio
Automatically enabled when the channel pair is
configured for thermocouple sensors.
The maximum open detection time is
3 seconds.
Channel configurations
Table 2. Channel configuration specifications
J, K, S, R, B, E, T, or N
Note 1: Channel configuration information is stored in the EEPROM of the isolated microcontroller by the
firmware whenever any item is modified. Modification is performed by commands issued over USB or
wireless from an external application, and the configuration is made non-volatile through the use of the
EEPROM.
Note 2: The factory default configuration is Type J.