WLS-TEMP Wireless Sensor Measurement User's Guide Document Revision 5 March 2013 © Copyright 2013

Seite 2 - HM WLS-TEMP.docx

WLS-TEMP Specifications 10 You can configure sensor settings when the WLS-TEMP is connected locally to the computer through the USB port, or when th

Seite 3 - Table of Contents

WLS-TEMP Specifications 11 Restoring factory default settings You can restore the factory default configuration settings with InstaCal. Connecting t

Seite 4 - Chapter 5

12 Chapter 3 Sensor Connections The WLS-TEMP supports the following temperature sensor types:  Thermocouple – types J, K, R, S, T, N, E, and B  R

Seite 5 - About this User’s Guide

WLS-TEMP Specifications 13 Do not connect two different sensor categories to the same channel pair The WLS-TEMP provides a 24-bit A/D converter for

Seite 6 - Introducing the WLS-TEMP

WLS-TEMP Specifications 14 Wiring configuration Connect the thermocouple to the WLS-TEMP using a differential configuration, as shown in Figure 3.

Seite 7 - WLS-TEMP block diagram

WLS-TEMP Specifications 15 Two-wire configuration The easiest way to connect an RTD sensor or thermistor to the WLS-TEMP is with a two-wire configur

Seite 8 - Installing the WLS-TEMP

WLS-TEMP Specifications 16 Three-wire configuration A three-wire configuration compensates for lead-wire resistance by using a single-voltage sense

Seite 9 - Configuring the hardware

WLS-TEMP Specifications 17 Four-wire, two-sensor A four-wire, two-sensor measurement configuration is shown in Figure 8. Figure 8. Four-wire, two R

Seite 10

WLS-TEMP Specifications 18 Figure 10. Schematic showing switch detection by digital channel DIO0 Caution! All ground pins on the WLS-TEMP (pins 9,

Seite 11 - Warm up time

19 Chapter 4 Functional Details Thermocouple measurements A thermocouple consists of two dissimilar metals that are joined together at one end. When

Seite 12 - Sensor Connections

2 HM WLS-TEMP.docx Your new Measurement Computing product comes with a fantastic extra — Management committed to your satisfaction! Thank you for cho

Seite 13 - Thermocouple connections

WLS-TEMP Specifications 20 error voltage = resistance of the thermocouple x 105 nA To reduce the error, reduce the length of the thermocouple to low

Seite 14 - Wiring configuration

WLS-TEMP Specifications 21 1 Screw terminal pins 1 to 26 6 Receive LED 2 Screw terminal pins 27 to 52 7 Transmit LED 3 Command LED 8 Wireless Power

Seite 15 - Two-wire configuration

WLS-TEMP Specifications 22 LED Function Received Signal Strength (RSS) Indicator LEDs 3 green LED bar graph. The LEDs will turn on when receiving a

Seite 16 - Four-wire configuration

23 Chapter 5 Specifications All specifications are subject to change without notice. Typical for 25 °C unless otherwise specified. Specifications in

Seite 17 - Digital I/O connections

WLS-TEMP Specifications 24 Channel configurations Table 2. Channel configuration specifications Sensor Category Condition Max number of sensors (all

Seite 18 - WLS-TEMP Specifications

WLS-TEMP Specifications 25 Accuracy Thermocouple measurement accuracy Table 4. Thermocouple accuracy specifications, including CJC measurement error

Seite 19 - Functional Details

WLS-TEMP Specifications 26 RTD measurement accuracy Table 6. RTD measurement accuracy specifications RTD Sensor Temperature (°C) Maximum Accuracy Er

Seite 20 - External components

WLS-TEMP Specifications 27 Note 12: Resistance values greater than 180 kΩ cannot be measured by the device in the thermistor mode. The 180 kΩ resi

Seite 21 - Status LEDs

WLS-TEMP Specifications 28 Temperature alarms Table 11. Temperature alarm specifications Parameter Specification Number of alarms 8 (one per digital

Seite 22 - LED Test button

WLS-TEMP Specifications 29 Note 16: Contains FCC ID: OUR-XBEEPRO. The enclosed device complies with Part 15 of the FCC Rules. Operation is subject

Seite 23 - Specifications

3 Table of Contents Preface About this User’s Guide ...

Seite 24 - Compatible sensors

WLS-TEMP Specifications 30 USB specifications Table 17. USB specifications Parameter Specification USB device type USB 2.0 (full-speed) Device compa

Seite 25 - Accuracy

WLS-TEMP Specifications 31 LED / button configuration Table 21. LED configuration Parameter Specification Command LED Green LED – indicates a comman

Seite 26 - RTD measurement accuracy

WLS-TEMP Specifications 32 Screw terminal connector Table 22. Screw terminal connector specifications Connector type Screw terminal Wire gauge range

Seite 27 - Digital input/output

Declaration of Conformity Manufacturer: Measurement Computing Corporation Address: 10 Commerce Way Suite 1008 Norton, MA 02766 USA Category

Seite 28 - Wireless communications

Measurement Computing Corporation 10 Commerce Way Suite 1008 Norton, Massachusetts 02766 (508) 946-5100 Fax: (508) 946-9500 E-mail: [email protected]

Seite 29 - USB +5V voltage

WLS-TEMP Specifications 4 Open-thermocouple detection (OTD) ...

Seite 30 - Mechanical

5 Preface About this User’s Guide What you will learn from this user's guide This user's guide describes the Measurement Computing WLS-TEMP

Seite 31 - LED / button configuration

6 Chapter 1 Introducing the WLS-TEMP Overview: WLS-TEMP features This user's guide contains all of the information you need to configure the WLS

Seite 32 - Screw terminal connector

WLS-TEMP Specifications 7 WLS-TEMP block diagram WLS-TEMP functions are illustrated in the block diagram shown here. Figure 1. Functional block dia

Seite 33 - Declaration of Conformity

8 Chapter 2 Installing the WLS-TEMP What comes with your shipment? The following items are shipped with the WLS-TEMP: Hardware  WLS-TEMP  AC-to-

Seite 34