Measurement-computing USB-5203 Bedienungsanleitung Seite 24

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 33
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 23
USB-5203 User's Guide Specifications
24
Table 4. Thermocouple accuracy specifications, including CJC measurement error
Sensor Type Maximum error (°C) Typical error (°C) Temperature range (°C)
J ±1.499 ±0.507 210 to 0
±0.643 ±0.312 0 to 1200
K ±1.761 ±0.538 210 to 0
±0.691 ±0.345 0 to 1372
S ±2.491 ±0.648 50 to 250
±1.841 ±0.399 250 to 1768.1
R ±2.653 ±0.650 50 to 250
±1.070 ±0.358 250 to 1768.1
B ±1.779 ±0.581 250 to 700
±0.912 ±0.369 700 to 1820
E ±1.471 ±0.462 200 to 0
±0.639 ±0.245 0 to 1000
T ±1.717 ±0.514 200 to 0
±0.713 ±0.256 0 to 600
N ±1.969 ±0.502 200 to 0
±0.769 ±0.272 0 to 1300
Semiconductor sensor measurement accuracy
The error shown in Table 5 below does not include errors of the sensor itself. These specs are for one year while
operation of the device is between 15 °C and 35 °C. Please contact your sensor supplier for details on the actual
sensor error limitations.
Table 5. Semiconductor sensor accuracy specifications
Sensor Type Temperature Range (°C) Maximum Accuracy Error (°C)
LM35, TMP35 or equivalent 40 to 150 ±0.50 (Note 2)
RTD measurement accuracy
The error shown in Table 6 below does not include errors of the sensor itself. The sensor linearization is
performed using a Callendar-Van Dusen linearization algorithm. These specs are for one year while operating
the device between 15 °C and 35 °C. The specification does not include lead resistance errors for 2-wire RTD
connections. Please contact your sensor supplier for details on the actual sensor error limitations.
In RTD mode, the device cannot measure resistance values greater than 660 Ω. The 660 Ω resistance limit
includes the total resistance across the current excitation (Ix+/Ix) pins, which is the sum of the RTD resistance
and the lead resistances.
For accurate three-wire compensation, the individual lead resistances connected to the Ix+/Ixpins must be of
equal value.
Table 6. RTD measurement accuracy specifications
RTD Sensor
Temperature (°C)
Maximum Accuracy Error (°C)
Ix+ = 210 µA
Typical Accuracy Error (°C)
Ix+ = 210 µA
PT100, DIN, US
or ITS-90
200 to 150 ±2.85 ±2.59
150 to 100 ±1.24 ±0.97
100 to 0 ±0.58 ±0.31
0 to 100 ±0.38 ±0.11
100 to 300 ±0.39 ±0.12
300 to 600 ±0.40 ±0.12
Seitenansicht 23