Measurement-computing USB-2416 Bedienungsanleitung Seite 21

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 49
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 20
21
Chapter 4
Calibrating the USB-2416
Measurement Computing calibrates the USB-2416 using an NIST-traceable calibration method. This method
stores a correction factor for each AIN range in non-volatile memory on the device at the time of calibration.
Allow a 45 minute warm-up period before calibrating the USB-2416.
Calibration methods
Factory calibration
Measurement Computing's Manufacturing Test department performs the initial factory calibration. You can
return the USB-2416 once a year for a new factory calibration and calibration certificate.
Self calibration
Use InstaCal to calibrate the USB-2416 whenever the ambient temperature changes by more than ±10 °C from
the last self-calibration
You can perform this procedure with thermocouple and voltage inputs connected to the input terminal blocks.
Seitenansicht 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 48 49

Kommentare zu diesen Handbüchern

Keine Kommentare