Measurement-computing StrainBook/616 Bedienungsanleitung Seite 112

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 146
  • Inhaltsverzeichnis
  • FEHLERBEHEBUNG
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 111
Synchronizing Multiple StrainBooks
Up to four StrainBook/616 units can be synchronized to each other via their SYNC ports. Each unit has
two identical SYNC ports. Either or both SYNC ports can be used to connect to the units via SYNC cables
CA-74-1 or CA-74-5 (1-foot or 5-foot cables, respectively). The units can be scan-synchronized and
triggered from any other SYNC-connected unit.
SYNC Cables
CA-74-1 (1 ft.) or
CA-74-5 (5 ft.)
Ethernet Cable
to PC
TR-40U Power
Adapter
Three StrainBook/616 Units, Synchronized
Not all trigger modes are supported in multiple StrainBook/616 Systems.
Synchronization Concept Block Diagram
The preceding diagram shows how a StrainBook/616 can input or output synchronization timing and
trigger events on either SYNC connector.
SYNC cables are not to exceed a total combined length of 15 feet (4.57 m).
10-8 System Expansion 968594 StrainBook/616 User’s Manual
Seitenansicht 111
1 2 ... 107 108 109 110 111 112 113 114 115 116 117 ... 145 146

Kommentare zu diesen Handbüchern

Keine Kommentare