Measurement-computing PCI-DAS6023 Bedienungsanleitung Seite 34

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 44
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 33
PCI-DAS6023 and PCI-DAS6025 User's Guide Specifications
34
Analog outputs (PCI-DAS6025 only)
Table 9. Analog output specifications (PCI-DAS6025)
D/A converter type
Double-buffered, multiplying
Resolution
12-bits, 1-in-4096
Number of channels
2 voltage output
Voltage range
±10 V
Monotonicity
12-bits, guaranteed monotonic
DNL
±1.0 LSB max
Slew rate
10 V/µs min
Settling time
20 V step to 0.012% (0.5 LSB): 10 µs max
Noise
200 µVrms, DC to 1 MHz BW
Glitch energy
±24 mV @ 2 µs duration measured at mid-scale transition.
Current drive
±5 mA
Output short-circuit duration
Indefinite @25 mA
Output coupling
DC
Output impedance
0.1 ohms max
Power up and reset
DACs cleared to 0 volts ±200 mV max
Table 10. Absolute accuracy specifications
Range
Absolute Accuracy
±10 V
±1.7 LSB
Table 11. Absolute accuracy components
Range
% of Reading
Offset
(mV)
Temp Drift
(%/DegC)
Absolute Accuracy at FS (mV)
±10 V
±0.0219
±5.93
±0.0005
±8.127
Each PCI-DAS6025 is tested at the factory to assure the board’s overall error does not exceed ±1.7 LSB.
Table 12. Relative accuracy specifications
Range
Relative Accuracy
±10 V
±0.3 LSB, typical
±0.5 LSB, max
Relative accuracy is defined as the measured deviation from a straight line drawn between measured endpoints
of the transfer function.
Seitenansicht 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 43 44

Kommentare zu diesen Handbüchern

Keine Kommentare