Appendix A Micro488/EX Command Summary
A.5
ONDOMACRO ON <event> DO[MACRO][number]
Enable Macro buffer
number
execution on <
event
>.
PASS
CONTROL PA[SS
CONTROL] addr
Make another bus device the Active Controller. Enter the
Peripheral (*CA) state.
PPOLL PPOLL
Read the Parallel Poll response from all bus devices.
PPOLL
CONFIG PPOLL C[ONFIG] addr,response
or
PPC addr,response
Set the Parallel Poll response of a bus device.
response
is the
decimal equivalent of four bits:
S
P2
P1
P0
.
PPOLL
DISABLE PPOLL
D[ISABLE]addr[,addr]
or PPD
addr[,addr]
Prevent bus device(s) from responding to a Parallel Poll.
PPOLL
UNCONFIG PPOLL
U[NCONFIG]
or
PPU
Prevent all devices from responding to a Parallel Poll.
READ READ[;][number]
Send a copy of the specified
number
Macro buffer out to the
serial host.
READ
LOG READ LOG
Send a copy of the
LOG
buffer out to the serial host.