Section 2 Getting Started
2.9
2.3.3.5 Trigger
(Pin 43)
The
Trigger
output is pulse for approximately 50
microseconds after a
GET
(Group Execute Trigger) command is
received from the bus controller. The trigger pulse is normally active
high, but can be made active low with the
Invert
command (
I2
). The
Handshake
command (
H2
) can independently pulse the
Trigger
line,
independent of any bus activity.
Timing Diagram for Trigger Output
Trigger
75 µS typ 50 µS typ
GET
2.3.3.6 Service
(Pin 45)
The
Service
input is an edge sensitive input capable of
generating a bus
Service Request
(
SRQ
)
.
It is enabled with the
SRQ
command (
M1
) and defaults to rising-edge sensitive. The
Invert
command (
I64
) can be used to program it to be falling-edge sensitive.
2.4 IEEE 488 Bus Implementation
The
Digital488
implements many of the capabilities defined by the IEEE 488
1978 specification. These are discussed in the following sections. Those bus uniline
and multiline commands that the
Digital488
does not support or respond to include:
Remote Enable
(REN)
Parallel Poll
(PP)
Go to Local
(GTL)
Parallel Poll Configure
(PPC)
Local Lockout
(LLO)
Parallel Poll Unconfigure
(PPU)
Take Control
(TCT)
Parallel Poll Disable
(PPD)