Section 2 Getting Started
2.3
2.2.2 IEEE 488 Bus Output Terminator Selection
The terminating characters sent on output by the
Digital488
are
determined by
SW1-6
through
SW1-8.
The terminator switches are read
only at power on, but can be changed by the controller through the
Terminator
command. If power is cycled after receipt of the
Terminator
command, then the unit will once again default to the switch settings. The
factory default settings are Carriage Return - Line Feed with EOI asserted.
The
Digital488
ignores all terminators received from the bus
controller. Only the
Execute
command
(X)
is used to signal the
Digital488
that a command string has been completed.
SW1 View for Terminator Selection
OPEN
12345678
1
0
DOT
Switch
Side
View
OPEN
12345678
1
0
OPEN
12345678
1
0
OPEN
12345678
1
0
CR-LF LF-CR
CR Only LF Only
OPEN
12345678
1
0
OPEN
12345678
1
0
EOI Enabled EOI Disabled