DAQFlex Message-based Firmware Specification Analog input calibration messages
15
Set the slope in hexadecimal bytes for the current AICAL:RANGE setting.
"AICAL{ch}:SLOPE/HEX=hex value"
The calibration slope in hexadecimal bytes.
"AICAL{0}:SLOPE/HEX=0x14B525A3"
The hex value is used to preserve the precision of the double precision floating point
value.
Get the slope in hexadecimal bytes for the current AICAL:RANGE setting.
"?AICAL{ch}:SLOPE/HEX"
or
"?AICAL{ch}:SLOPE"
Refer to the device-specific topics to see whether "/HEX" is supported.
"AICAL{ch}:SLOPE=0xhex value"
The calibration slope in hexadecimal bytes.
TCCAL
Enables or disables the TC Cal Measurement option.
Get the value for the TC Cal Measurement option.
UNLOCK
Unlock the AICAL component.
The AICAL component must be unlocked in order to change calibration settings on a
device's analog inputs.
When AICAL is unlocked, sending a message to any component other than AICAL will
lock the AICAL component.