Measurement-computing E-1608-OEM Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Hardware Measurement-computing E-1608-OEM herunter. Measurement Computing E-1608-OEM User Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - User's Guide

E-1608-OEM Ethernet-based High-speed Multifunction DAQ Document Revision 1 December 2014 © Copyright 2014 User's Guide

Seite 2 - HM E-1608-OEM.docx

E-1608-OEM User's Guide Installing the E-1608-OEM 10 Setting up the E-1608-OEM for communication across networks In order to communicate with th

Seite 3 - Table of Contents

E-1608-OEM User's Guide Installing the E-1608-OEM 11 Updating firmware Your DAQ device contains firmware that can be updated in the field if req

Seite 4

12 Chapter 3 Functional Details Analog input modes The E-1608-OEM can acquire analog input data in two basic modes – software paced and hardware pace

Seite 5 - About this User's Guide

E-1608-OEM User's Guide Functional Details 13 Ethernet connector The E-1608-OEM has one 10/100 BASE-T, auto-negotiation, high-speed communicatio

Seite 6 - Introducing the E-1608-OEM

E-1608-OEM User's Guide Functional Details 14 Connector W2 Header connector W2 provides connections for the DIO, external clock I/O, trigger, co

Seite 7 - Functional block diagram

E-1608-OEM User's Guide Functional Details 15 Leave unused input channels either floating or tied to an AGND connector. Source impedances should

Seite 8 - Installing the E-1608-OEM

E-1608-OEM User's Guide Functional Details 16 Ethernet to the same local network. Typical throughput is not guaranteed if a wireless connection

Seite 9 - Configuring network settings

E-1608-OEM User's Guide Functional Details 17 Figure 6. W3 jumper configurations For more information about digital signal connections For gene

Seite 10 - Calibrating the hardware

E-1608-OEM User's Guide Functional Details 18 Mechanical drawings Figure 7. E-1608-OEM dimensions

Seite 11 - Updating firmware

Chapter 4 Specifications All specifications are subject to change without notice. Typical for 25 °C unless otherwise specified. Specifications in it

Seite 12 - Functional Details

HM E-1608-OEM.docx Your new Measurement Computing product comes with a fantastic extra — Management committed to your satisfaction! Thank you for

Seite 13 - Header connectors

20 Accuracy Analog input DC voltage measurement accuracy Table 2. DC Accuracy components and specifications. All values are (±) Range Gain error (% o

Seite 14 - Signal connections

21 Analog output Table 5. Analog output specifications Parameter Condition Specification Number of channels 2 Resolution 16 bits Output ranges Cali

Seite 15 - Analog output

22 Analog input/output calibration Table 9. Analog input/output calibration specifications Parameter Specification Recommended warm-up time 15 minute

Seite 16 - Digital I/O

23 External trigger Table 11. External trigger specifications Parameter Condition Specification Trigger source External digital TRIG Trigger mode Sof

Seite 17 - Power output

24 Counter Table 13. Counter specifications Parameter Specification Pin name CTR Counter type Event counter Number of channels 1 Input type Schmitt t

Seite 18 - Mechanical drawings

25 Network Ethernet connection Table 16. Ethernet connection specifications Parameter Specification Ethernet type 100 Base-TX 10 Base-T Communication

Seite 19 - Specifications

26 LED displays and the factory reset button Table 20. LED and button configurations Parameter Specification Power LED (top) 3.3 V < Vext < 5.9

Seite 20 - Accuracy

27 Table 24. W2 pinout Pin Signal name Pin description Pin Signal name Pin description 1 DIO0 DIO channel 0 2 DIO1 DIO channel 1 3 DIO2 DIO channel 2

Seite 22 - Digital input/output

Measurement Computing Corporation 10 Commerce Way Suite 1008 Norton, Massachusetts 02766 (508) 946-5100 Fax: (508) 946-9500 E-mail: [email protected]

Seite 23 - External clock input/output

3 Table of Contents Preface About this User's Guide ...

Seite 24 - Counter

E-1608-OEM User's Guide 4 Analog input DC voltage measurement accuracy ...

Seite 25 - Network

5 Preface About this User's Guide What you will learn from this user's guide This user's guide describes the Measurement Computing E-1

Seite 26 - Mechanical

6 Chapter 1 Introducing the E-1608-OEM The E-1608-OEM is compatible with both TCP/IP (IPv4 only) and user datagram protocol (UDP) network protocols.

Seite 27

E-1608-OEM User's Guide Introducing the E-1608-OEM 7 Functional block diagram E-1608-OEM functions are illustrated in the block diagram shown he

Seite 28

8 Chapter 2 Installing the E-1608-OEM Unpacking As with any electronic device, you should take care while handling to avoid damage from static elect

Seite 29

E-1608-OEM User's Guide Installing the E-1608-OEM 9 Configuring network settings The following E-1608-OEM network settings are software-selectab