WaveCal, pg. 2
05-22-01
Last Cal displays the date the system was last calibrated. The date indicates when calibration constants
were last written to WaveBooks internal EEPROM.
;
Calibrate
: Checking the
Calibrate
check-box informs
WaveCal
that the applicable module and its
related options are to be calibrated for at least one channel.
;
Concurrent
: Checking the
Concurrent
check-box informs
WaveCal
that all module channels are to
be calibrated simultaneously, instead of individually. If this box is not checked the 2-point calibration
process calibrates each channel (of the applicable module) on an individual basis.
By using the Concurrent method you will greatly reduce your system’s calibration time.
Make sure all of the system modules and options are properly connected and correctly listed in the
WaveBook Calibration Tool
window. After this verification you may select the module(s) to calibrate and
whether their channels are to be individually or concurrently calibrated.
Performing System Calibration
Open the Calibrate pull-down menu and click
Start
to access a
User Calibration
box
for the selected module (see following figure).
Aspects of the
User Calibration
dialog box are discussed on the following page.
The Calibrate
Pull-Down Menu
WaveCals WaveBook Calibration Tool and User Calibration Box