PCI-DAS6013 and PCI-DAS6014 User's Guide Specifications
32
Table 2. Absolute accuracy components specifications - All values are (±)
Averaged Noise +
Quantization (µV)
1
Absolute
Accuracy at FS
(mV)
1
Averaged measurements assume averaging of 100 single-channel readings.
Each PCI-DAS6013 and PCI-DAS6014 is tested at the factory to assure the board’s overall error does not
exceed accuracy limits described in Table 1.
Table 3. Differential non-linearity specifications
±0.5 LSB ty,p ±1.0 LSB max
System throughput
1. Single channel, single input range
2. Multiple channel, single input range
3. Single channel, multiple input ranges
Default to value for cbAInScan() range parameter
Note 2: For conditions 1-2 above, specified accuracy is maintained at rated throughput. Condition 3
applies a calibration coefficient which corresponds to the range value selected in cbAInScan().
This coefficient remains unchanged throughout the scan. Increased settling times may occur
during gain-switching operations.
Settling time
Settling time is defined as the time required for a channel to settle to within a specified accuracy in response to
a full-scale (FS) step. Two channels are scanned at the specified rate. A –FS DC signal is presented to
channel 1; a +FS DC signal is presented to channel 0.